e-Series Nitrogen Oxides Analyzer NO, NOx & NO2

AC32e

CLD based gas monitor offering superior performances for the measurement of NO, NO2 and Nox in the range 0-1 ppm or 0-10 ppm

Eco-designed, reference air quality monitor, based on the chemiluminescence technology, the standard method for the measurement of nitrogen oxides (EN 14211). QAL1 certified by the TÜV, US-EPA approved. With its ultra low power consumption & 24 V supply (option), the AC32e is the sole NOx analyzer on the market able to measure without being connected to the power grid.

  • Excellent sensitivity & metrological performances for NO, NO2 and NOx measurements in the range 0-1 ppm
  • Real-time calibration graph, animated synoptic, auto-diagnostic, control and maintenance data screens can be displayed while the instrument is operating
  • Service assistance inside: detects early signs of trouble, allows predictive maintenance, identifies the needed service and guides service operations step by step: increased productivity on site, reduced downtime, more efficiency, less training…
  • Embedded Communication Protocol for XR® Software with automatic recognition & configuration
  • Able to withstand temperatures up to 45-50 °C without air conditioning
  • Smart, connected instrument with embedded web server for the easiest, full remote control and operation of the analyzer from any PC, tablet or smartphone (configuration, calibration, test, diagnostic parameters…)
  • Continuous indoor and outdoor air quality monitoring
  • Stationary and mobile AQMS laboratories
  • Industrial fence-line monitoring
  • Continuous emissions monitoring (CEM) by dilution
  • Background (urban or suburban), Rural, Traffic, Kerbside measurement campaigns and monitoring studies
  • Laboratory and field studies on pollution effects
Technical specifications
Measured parameter NO, NO₂, NOx
Technology Chemiluminescence
US-EPA Approval / QAL 1 Certification YES
Measurement ranges 0-1 or 0-10 ppm
Measurement units ppb, ppm, µg/m3 (user-selectable)
Lower Detectable Limit (2σ) < 0.2 ppb
Noise (σ) 0.1 ppb
Zero drift < 1 ppb / 24h
Span drift < 1% / 7 days
Linearity 1%
Sample flow rate 0.66 l/min (1 l/min with dryer)
Sampling pump External
Temperature range 0°C to +40°C
Data storage 1 year
Included I/O 1LAN/3USB/WIFI
Weight (Kg) 10.3 (+ 4,6 pompe)
Dimensions mm (LxWxH) 483 x 606 x 133
Chassis 19″ Rack, 3U
Power supply ‘100~250Vac, 50/60Hz + ground
    MAIN OPTIONS:

  • TFT color touch screen (800 x 480 , 7″)
  • Additional parameter: NH3
  • 24 V Power supply PLUS extended temperature range 0-50°C

Our monitoring solutions are European QAL 1 certified and in compliance with the latest international regulations & standards.
They are also approved and certified by various laboratories and organizations around the world such as: US EPA, TÜV, GOST, CEN, JQA, CNEMC, JMOE, KTL, CNSA, LCSQA…

    COMPLIANCE WITH:

  • ISO 7996, EN 14211 (2012), EN 15267 (2009), 40 cfr part 53 & part 58
  • QAL 1 certified by the TÜV to N° 0000053805
  • U.S. EPA approved as compliant with RFNA-0118-249

Related products

QUOTE CONTACT

Compare up to 5 products Show

Le nouveau site d’ENVEA est actuellement en cours de développement, fermez cette fenêtre pour poursuivre votre navigation. Vous pouvez aussi retrouver l’intégralité des rubriques en cliquant sur le bouton ci-dessous.

The new ENVEA website is currently under development, close this window to proceed your navigation. You can also find all the sections by clicking on the button below.
ENVEA
close-link
Previous Next
Close
Test Caption
Test Description goes like this
This site uses cookies. By the further use you hereby agree. For detailed information, please refer to the Cookies. Ok