- Heated analyzer (temperature controlled up to 180°C) with measuring chamber under vacuum minimizing the quenching effect
- Designed to measure wet & corrosive samples
- Automatic CO2 and H2O quenching correction
- Proven EPA reference method
- CLD technique for continuous and simultaneous measurement of NO, NO2 and NOx, in compliance with EN 14792
- Accurate measurement in the range: 0-10/100/1 000/10 000 ppm
- Graphic Liquid Crystal Display (LCD) with interactive menu driven software with enhanced speed display
- Built-in memory for data storage
- Built-in Ethernet TCP/IP connection for remote control and serial interface RS232
- Fast response time (<2 sec)
- Communication by AK protocol (RS232 / RJ45)
- Available in 2 versions:
TOPAZE 52M-S : NO or NOx monitoring
TOPAZE 52M-D : NO-NOx and NO2 simultaneous monitoring
- Engine manufacturing and testing in the automotive, heavy duty, aviation, marine, rail transport sectors, etc.
- Continuous Emission Monitoring
- Combustion control
- Process monitoring
- Laboratories and Research Centres…
Measurable parameters | |||
---|---|---|---|
Pollutant | Range (ppm) |
||
NO | 0-10 / 10000 | ||
NO2 | 0-10 / 10000 | ||
NOx | 0-10 / 10000 |
Lowest / Highest available ranges expressed in mg/m3 (may vary with your site conditions to be indicated on the Site Survey Form)
Technical Specifications | |
---|---|
QAL1 to EN 15267-3 certified | NO |
US EPA compliance | YES |
Sampling technology | Hot/Wet Extraction |
Sample transfert | 180°C heated sampling line |
back flush function | YES |
span injection | YES on analyser or at sample point |
Built-In O2 option | NO |
Serial link communication | RS232 / RS422 |
Other Communication | Ethernet |
Analogue OUTPUTS | Optional |
Analogue INPUTS | NO |
Dimensions (mm) | 19” rack: 483 x 440 x 135 (L x W x H) |