Continuous, Automatic Suspended Particulate Monitor

MP101M

PM10, PM2.5 and PM1 fine dust concentration measurement in the range 0-10,000 μg/m3 using the Standard Reference Method

The MP101M analyzer uses the standard ISO 10473 beta gauge measurement method for the continuous measurement of the concentration of fine dust in ambient air. The instrument is certified equivalent to the reference method in accordance with the EN12341 European standards for PM10 and PM2.5 respectively EN14907 and US EPA for the continuous monitoring of PM10 or PM2.5.

  • High accuracy, not influenced by the physicochemical nature, color or shape of particles
  • Fibreglass tape with 3 years of autonomy of continuous daily sampling
  • Rugged instrument, insensitive to vibration, humidity, temperature…
  • Built-in reference gauge for calibration: no need for factory recalibration
  • Sampling flow-rate continuously regulated to the atmospheric temperature and pressure: reduces evaporation artefacts of volatile compounds (mandatory for PM2.5 according to EU regulations)
  • Equipped with large, 7’’ colour touchscreen with animated, real-time display
  • Interchangeable sampling heads for PM10, PM2.5, PM1 and TSP
  • Smart, connected instrument with embedded web server for the easiest, full remote control and operation of the analyzer from any PC, tablet or smartphone (configuration, calibration, test, diagnostic parameters…)
  • Cost-saving analyzer with ultra low power consumption, 24 V supply (option) and able to withstand temperatures up to 45-50 °C without air conditioning
(*) Light scattering technologies applied to particle mass concentration measure can be affected by aerosols chemical composition and atmospheric conditions & should be subject to operator interpretation.
  • Urban, industrial, background… air pollution monitoring networks
  • Mobile air quality monitoring laboratories
  • Roadside air quality monitoring, traffic pollution control
  • Ambient Air Pollution Hotspot Monitoring
  • Process leakage, fence line monitoring for industry
  • Air pollution Measurement Campaigns and Monitoring Studies
Technical specifications
Measured parameter PM10, PM2.5, PM1, TSP
Technology Beta Ray Attenuation
US-EPA Approval / QAL 1 Certification YES
Measurement ranges up to 10 000 μg/m3
Measurement units µg/m3
Lower Detectable Limit (2σ) 0.5 μg/m3 (24h average)
Sample flow rate 1 m³/h
Sampling pump External (vacuum pump)
Temperature range +5°C to 40°C
Data storage Unlimited
Included I/O LAN/USB/WIFI/AO
Weight (Kg) 15.2 (+pump: Picolino 4.6Kg / KNF 9.4Kg)
Dimensions mm (LxWxH) 360.5 x 483 x 266
Chassis 19″ Rack, 3U
    MAIN OPTIONS:

  • TFT color touch screen (800 x 480 , 7″)
  • 24 V Power supply PLUS extended temperature range 0-50°C
    The instrument is certified equivalent to the reference method in accordance with the EN12341 European standards for PM10 and PM2.5 respectively EN14907 and US EPA for the continuous monitoring of PM10 or PM2.5.

  • STANDARD COMPLIANCE ISO 10473: 2000
    FOR PM10 :
    US-EPA (EQPM-0404-151)
    EN 12341 (I-CNR 087/2004, F-LCSQA)
    FOR PM2.5:
    EN 14907 (F-LCSQA)
    US-EPA (EQPM-1013-211)
    J-MOE PM 2.5 TYPE APPROVED

Related products

QUOTE CONTACT

Compare up to 5 products Show

Le nouveau site d’ENVEA est actuellement en cours de développement, fermez cette fenêtre pour poursuivre votre navigation. Vous pouvez aussi retrouver l’intégralité des rubriques en cliquant sur le bouton ci-dessous.

The new ENVEA website is currently under development, close this window to proceed your navigation. You can also find all the sections by clicking on the button below.
ENVEA
close-link
Previous Next
Close
Test Caption
Test Description goes like this
This site uses cookies. By the further use you hereby agree. For detailed information, please refer to the Cookies. Ok