Generation of mercury calibration gas
The MC-3000 generates a continuous stream of a mercury vapor loaded gas in order to check or calibrate mercury analysers.
An exactly adjustable carrier gas stream (normally air or nitrogen) is loaded with mercury in a first step.
In a second step the gas is supersaturated by cooling. Subsequently the excess mercury is condensed in a special cell to get exact saturation. The gas is then diluted, the dilution ratio is continuously adjustable in order to obtain concentrations within a specific range.
Approx. 10 to 500 µg / m3 Hg°
1 to 9 l/min, depending on conc.
Air, Nitrogen
Max. 12.5 l(min)
1 to 3 bar
Built-in particular filter (0.2 µm )
Sulphurized activated carbon at outlet
Pt-100 with 0.01 °C accuracy
Electronic mass flow controller (MFC)
For tubing 4mm i. d. / 6mm o. d.
230 VAC / 50 Hz (110 VAC / 60 Hz option)
30 VA
45 x 18 x 27 cm
approx. 5 kg